ACCU-TELL FOB CASSETTE TEST (CE)

Declaration: As with all diagnostic tests, a definitive clinical diagnosis should not be based on the result of a single test, but should only be made by the physician after all clinical and laboratory findings have been evaluated.

 

CATALOG

INTENDED USE

PRECAUTIONS

SPECIMEN COLLECTION AND PREPARATION

                                                                                PERFORMANCE CHARACTERISTICS
                                                                                REFERENCE

 

CATALOG

Catalog No.                    Product Name                    Note

ABT-FOB-B22                 FOB Cassette                    CE

 

INTENDED USE
Accu-Tell ®  Rapid FOB Test is a rapid, visual immunochromatographic test for the qualitative detection of human blood hemoglobin in fecal samples. This test is intended as an aid in the diagnosis of lower gastrointestinal (g.i.) disorders. The test is recommended for professional use. For in vitro diagnostic use only.
The principal use of FOB test is to screen for lower g.i. pathologies, such as colorectal cancers and large adenomas that bleed. Colorectal cancer is one of the most commonly diagnosed cancers and a leading cause of cancer death in the United States (Lieberman, 1994; MMWP, 1995). Screening for colorectal cancer probably increases the cancer detection at an early stage, therefore reduces the mortality (Dam et. al., 1995; Miller, 1995; and Lang, 1996). Earlier commercially available FOB tests utilized the guaiac test, which requires special dietary restriction to minimize false positive and false negative results. Rapid FOB Test are specially designed to detect human hemoglobin in fecal samples using Immunochemical methods, which improved specificity for the detection of lower g.i. disorders, including colorectal cancers and adenomas (Frommer et. al., 1988; St. John et. al., 1993).
PRINCIPLE
Accu-Tell ®  Rapid FOB Test has been designed to detect human hemoglobin in fecal samples through visual interpretation of color development in the test device. The test device contains a membrane strip, which is pre-coated with anti-human hemoglobin antibody on the Test Line region (T) and goat anti-mouse antibody on the Control Line region (C). An anti-human hemoglobin antibody-colloidal gold conjugate pad is placed at the end of the membrane. When human hemoglobin is present in the patient fecal sample dissolved in buffered saline, the mixture of colloidal gold conjugate and extracted sample moves along the membrane chromatographically by capillary action. This mixture then migrates to the Test region (T) and forms a visible line as the antibodies complex with the human hemoglobin. When human hemoglobin is absent in the extracted sample, no visible color band will form on the Test region (T). Therefore, the presence of a color band in the Test region (T) indicates a positive result. A colored band will always appear at the Control region (C) to serve as a procedural indicator for the proper performance of the test and the device.
STORAGE AND STABILITY
The test kit is to be stored at refrigerated (2-8°C) or at room temperature (up to 30°C) in the sealed pouch for the duration of the shelf-life.
PRECAUTIONS
1.  For in vitro diagnostic use only.
2.  For professional use.
3.  Do not use test kit beyond expiration date.
4.  Do not mix Sample Collection Tubes from different lots.
5.  Do not open the Test Cassette foil pouch until you are ready to perform the test.
6.  All patient samples should be treated as if capable of transmitting disease.
7.  Buffered Saline contains sodium azide which may react with lead or copper plumbing to form potentially explosive metal azides. When disposing of buffered saline or extracted samples, always flush with copious quantities of water to prevent azide build up.
8.  Patients should closely follow the specimen collection procedures. Patients should not collect samples during their menstrual period, if they have bleeding hemorrhoids, blood in the urine, or if they have strained during bowel movement.
SPECIMEN COLLECTION AND PREPARATION
1.  Collect a random sample of feces in a clean dry container or receptacle.
2.  Unscrew and remove the collection tube applicator stick. Be careful not to spill or spatter solution from container.
3.  Collect random sample by using the applicator stick. Take sample from various surfaces of the feces specimen
4.  Re-insert the applicator stick into the tube and screw the cap tightly. Be careful not to break the tip of the sample collection tube.
5.  The specimen is now ready to be stored at 2-8°C, transported or tested. Fecal samples in the buffered saline are stable for up to 15 days at room temperature.

 

TEST PROCEDURE
Quality Controls
Internal Procedural Control
1.  A procedural control is included in the test. A colored band appearing on the Control region (C) of the membrane indicates proper performance of the test and the device.
2.  A clear background in the observation window is considered an internal negative control. However, when the fecal samples are tested, the background may appear slightly yellowish due to the original color of the fecal samples. This is acceptable as long as it does not interfere with the interpretation of test result. The test is invalid if the background fails to clear and obscures the reading of the result.
ASSAY PROCEDURE
1.  Test device, patient’s samples, (extracted sample) should be brought to room temperature (20°C to 30°C) prior to testing.
2.  Remove the test device from its pouch when ready to perform the test. Bring the device to room temperature to avoid condensation of moisture in the membrane. Label the device with patient or control identification.
3.  Shake the collection tube thoroughly to ensure proper mixing of the fecal sample with the extraction solution.
4.  Using a piece of tissue paper, break the tip of the collection tube using a twisting motion. Hold the collection tube vertically and dispense 4-6 drops of solution into the sample well of the test device.
5.  Observe the result in 5 minutes. Strong positive results may be observed sooner. Do not interpret ate after 8 minutes.

 

INTERPRETATION OF RESULTS
Positive: Two pink-red colored bands appear. One in the control region (C) and one in the test region (T). When testing with strong positive samples, the intensity of the control band may be lighter than expected. Comparison of the line intensities is not recommended.
Negative: Only one pink-red colored band appears in the control region (C). No apparent faint pink to red colored band in the test region (T).
Invalid: A total absence of pink colored bands in both regions is an indication of procedural error or that test reagents may have deteriorated. Repeat the test with a new test device

 

LIMITATIONS
1.  This test kit is to be used for the qualitative detection of human hemoglobin in fecal samples. A positive result suggests the presence of human hemoglobin in fecal samples. The presence of blood in stools may be due to several causes, besides colorectal bleeding, such as hemorrhoid, blood in urine or stomach irritations.
2.  All colorectal bleedings may not be due to precancerous or cancerous polyps. The data obtained by this test should be used in conjunction with other clinical findings and testing methods, such as barium enema, sigmoidoscopy or colonoscopy gathered by the physician.
3.  Negative results do not exclude bleeding since some polyps and colorectal region cancers can bleed intermittently or not at all. Additionally, blood may not be uniformly distributed in fecal samples. Colorectal polyps at an early stage may not bleed.
4.  Urine and excessive dilution of sample with water from toilet bowl may cause erroneous test results.
5.  This test may be less sensitive for detecting upper g.i. bleeding because blood degrades as it passes through the g.i. track.
PERFORMANCE CHARACTERISTICS
A. Analytical Sensitivity:
A sample containing human hemoglobin at concentration equal to or higher than 100 ng/ml produces a positive result. In some cases sample containing human hemoglobin at concentrations less than 100 ng/ml can also be tested as positive.
Prozone effect: sample containing as high as 0.5 mg/ml hemoglobin can still test positive.
B. Test Specificity:
Rapid FOB Test is specific for human hemoglobin and does not show any cross-reaction with the hemoglobin from bovine, pig, rabbit, horse and sheep concentrations up to 0.5 mg/ml.
Rapid FOB Test also does not show any cross reaction with billirubin, vitamin C and horse radish peroxidase.
REFERENCE
1.  Dam, J.V., et. al.; Feccal Blood Screening for Colorectal Cancer; Archive of Internal Medicine; (1995) 155:2389-2402.
2.  Frommer, D.J. et. al.; Improved Screening for Colorectal Cancer by Immunological Detection of Ocult Blood; British Medical Journal; (1988) 296:1092-1094.
3.  Lieberman, D.; Screening/Early Detection Model for Colorectal Cancer, Why Screeen? Cancer Supplement; (1994) 74(7):2023-2027.
4.  Miller, A.B.; An Epidemiological Perspective on Cancer Screening; Clinical Biochemistry (1995) 28(1): 41-48.
5.  Ransohoff, D.F. and Lang, C.A.; Improving the Feccal Occult-Blood Test; The New England Journal of Medicine; (1996) 334 (3):189-190.
6.  Screening for Colorectal Cancer-United States, 1992-1993, and New Guidelines; Mobility and Mortality Weekly Report; (1995) 45 (5): 107-110.

7.  St. John, D.J.B., et al.; Evaluation of New Occult Blood Test for Detection of Colorectal Neoplasia; Gastroenterology; (1993) 104:1661-1668.

 

The above information is just for reference. The actual technical specifications are subject to the insert provided with the product.